Measurement device and method of MTF for visible TV camera
Keywords:
Electro-optical imaging system, Line spread function, Modulation transfer functionAbstract
Modulation transfer function (MTF) is an important parameter to evaluate the imaging performance of TV cameras. A measuring device and method for MTF of visible TV camera were introduced. A threeintegral spherical target generator system was designed, the target and background brightness can be independently controlled by electric attenuators. The outlet non-uniformity is 4.1%. An automatic sub-pixel interpolation algorithm based on sliting slit method and a least square method of line spread functions were wtilized to realize the angle calculation automaticly and MTF measurement accurately. The repeatability of the measuring device is 0.007, and the measurement uncertainty is U=0.05 (k=2). Finally, the comparison experiment with foreign advanced instruments shows that, the measurement device established can effectively solve the problem of evaluating the image quality of TV cameras. © 2020, Editorial Board, Journal of Applied Optics. All right reserved.
References
Huang Y., Li X., Niu J., Error analysis of static MTF measurement for remote sensor in laboratory, Optics & Optoelectronic Technology, 15, 2, pp. 57-60, (2017);
Guan M., Zhang X., Zhang N., MTF measurement of CCD digital camera based on improved slanted-edge method, Journal of Applied Optics, 37, 1, pp. 100-103, (2016);
Han C., Performance optimization of electrooptical imaging systems, Optics and Precision Engineering, 23, 1, pp. 1-9, (2015); Fei X., Ni X., Present status and development trend of OTF testing instrument, Optical Instruments, 25, 2, pp. 48-52, (2003);
Zhu X., Zhang J., The characteristic of optical transfer function, Journal of Shaoyang University, 2, 2, pp. 18-20, (2005);
Tong W., Infrared system MTF technology analysis, Infrared, 4, pp. 16-22, (2003);
Bian J., Ma D., Sun G., MTF test technology anglysis of infrared electro-optical imaging system, Journal of Applied Optics, 34, 5, pp. 748-753, (2013);
Xu B., Shi Z., Feng B., Modulation transfer function measurement method of electro-optical imaging system, Acta Optica Sinica, 31, 11, (2011);
Yang H., The calibration of optical transfer function equipment of large aperture in infrared and visible band, Optical Technique, 1, pp. 87-94, (2001);
Ma W., Ma Z., He L., Et al., Influency of spatial frequency calibration on OTF test percision, Journal of Xi'an Institute of Technology, 24, 2, pp. 144-147, (2004);
Zhang M., Research on imaging evaluation of CCD camera basing on MTF method, (2015);
Yang H., Kang D., Jiang C., Et al., Visible light photoelectric imaging system machine integrated parameter calibration technology research, Journal of Applied Optics, 32, 2, pp. 254-258, (2015);
Cheng Y., On-orbit modulation transfer function(MTF) measurement for space camera, (2018); Li H., Yan C., Shao J., Et al., High accuracy measurement of the MTF of electro-optical imaging system based on modified slanted-edge method, Journal of Photons, 45, 12, (2016);
He X., Ji Q., Study on modulation transfer function measurement instrument with matrix for matrix CCD, Journal of Electronic Measurement and Instrumentation, 28, 10, pp. 1100-1108, (2014)
Published
How to Cite
Issue
Section
License
Copyright (c) 2020 Journal of Applied Optics
This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.
The CC Attribution-NonCommercial 4.0 License allows sharing and adapting the work, provided the creator is credited and the work is not used commercially. Modifications must be indicated, and derivative works under the same license are allowed.